Example 2.
In this example, the analyzer has been set up with the following parameters:
Start frequency = 90 MHz
Stop frequency 210 MHz
Number of points = 11
Maximum limit line begin frequency 90 MHz
Maximum limit line end frequency = 200 MHz
Refer to the illustration below for the discussion. Here, the data trace passes
through the limit line, and is above it, but this test passes because the last
point
is at 198 MHz. The last point (210 MHz) does not cause the limit
test to fail because it is past the limit line stop frequency of 200 MHz. But
there is an area on the data trace between 198 MHz and 200 MHz that is
above the displayed limit line.
E N D L I M I T
a t 2 0 0 M H z
M A X I M U M
DATA
L I M I T L I N E
I N Q U E S T I O N
N E X T
I
D A T A P O I N T
a t 2 1 0 M H z
l ’
I
I
A
.-.
9 0
1 0 2 1 1 4 1 2 6 1 3 8 1 5 0 1 6 2 1 7 4 1 8 6 1 9 8
D A T A P O I N T S
Figure 4-18. limit lines Example 2
4 - 4 9
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...