I
Using Instrument Functions
Using Markers
To Use Marker Flatness
The marker flatness search function measures a user-defined segment of the
measurement trace and calculates the following:
frequency span
gain
slope
flatness
The analyzer calculates flatness by drawing a straight line between the
markers. A maximum vertical deviation from this line is computed for each
measurement point. Flatness is the magnitude difference of the maximum
and minimum calculated deviations from the straight line.
Limit testing may be performed on the flatness parameter. See
Use
Marker Limit Functions,’ later in this chapter for information.
1. On measurement channel 1 press
and set markers 1 and 2
to define the beginning and end of the trace segment that you want to
measure. (When using measurement channel 2, use markers 3 and 4 to
define the trace segment.)
3. Figure 4-13 shows a
trace segment. Notice the marker readout in
the upper right corner of the display.
4-23
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...