Determine if a Calibration is Necessary
This section shows you how to determine if your measurement system
requires a
calibration.
l
Your test doesn’t require the best accuracy possible.
l
Your test device is connected directly to the reflection port with no adapters
or intervening cables.
l
Your test device impedance matches the impedance of the analyzer.
If your test setup meets these conditions, you do not need to perform any
additional calibrations, however without a user-calibration, the analyzer is not
guaranteed to meet its published measurement port specifications.
l
You want the best accuracy possible.
l
You are adapting to a different connector type or impedance.
l
You are connecting a cable between the test device and the analyzer test
ports.
l
You are measuring a narrowband or electrically long device.
l
You are connecting any attenuator or other such device on the input or
output of the test device.
6-7
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...