Using Instrument Functions
Using limit Testing
Peak-to-Peak Ripple
1. This limit test requires that you first define a segment on the measurement
trace using markers 1 and 2 (or markers 3 and 4 for measurement
channel 2). Then press [MARKER)
to enable the statistics marker search. See “lb Use Marker
Statistics” earlier in this chapter for more information on using marker
statistics.
2. Press
.
3. Use the front panel knob or the
keys to select Statistic: p-p in
the marker limit test table. Turn this limit function on by pressing the
OFF key. Note that the entry in the on/off column of the
table changes to “on.”
4. Press
and enter the maximum
limit using the front panel keypad and terminating the entry with the
k e y .
Press
Limit
and enter the minimum limit
using the front panel keypad and terminating the entry with the
key.
N O T E
Note that there are no visible limit lines or indicators on the display with this limit test function.
Therefore the
function has no effect when using marker limits.
4-40
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...