Optimizing Measurements
Increasing
Sweep Speed
Auto sweep time mode (the preset instrument mode), maintains the fastest
sweep speed possible for any particular measurement settings.
Press
and look at the
label.
When AUTO is all capital letters, it indicates that the analyzer is in auto
sweep time mode. If MAN is all capital letters, the analyzer is in manual
sweep time mode.
necessary, press
to toggle the time mode to
AUTO.
Wide system bandwidth is recommended for some broadband detection
measurements.
Press
to widen the IF’ bandwidth. As the
bandwidth increases, the sweep time decreases and the trace noise increases.
Table
Relationship Between System Bandwidth and Sweep Speed
Bandwidth
Wide
Medium Wide
Medium Narrow
Narrow
Fine
Speed
fastest
faster
fast
slow
slower
I slowest
5-4
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...