![horiba Fluorolog-3 Скачать руководство пользователя страница 86](http://html1.mh-extra.com/html/horiba/fluorolog-3/fluorolog-3_operation-manual_2162215086.webp)
Fluorolog-3 v. 2.2 (11 Jul 2002)
System Maintenance
5-19
To insert a new grating,
1
Place the 3 mounting pins on the back of the
grating into the matching mounting slots on the
grating stand.
2
Tighten the thumbscrew on the back of the
grating stand.
3
Lift the lid of the spectrometer off of the lid
mount.
4
Remove the lid mount.
5
Place the lid on the spectrometer.
6
Secure the lid with the Phillips screws.
No further adjustments or alignments are necessary.