Fluorolog-3 v. 2.2 (11 Jul 2002)
System Operation
3-8
6
Find the calibration peak for the xenon-lamp
spectrum.
The 467-nm peak is used for excitation calibration. The intensity of this
spectrum should be noted for reference, although it is not used for
instrument specifications.
Is the peak at
467 ± 0.5 nm?
Instrument is within
specification.
Go to emission cali-
bration check.
Instrument is
not
within
specification.
Recalibrate excitation
monochromator.
No
Yes