Fluorolog-3 v. 2.2 (10 Sep 2002)
Optimizing Data Acquisition
4-10
Note:
Measure the
G
fac-
tor before a polarization
experiment.
Measuring the
G
factor
The grating factor, or
G
factor, ought to be included
anytime polarization measurements are taken. The
G
factor corrects for variations in polarization
wavelength-response for the emission optics and
detectors. A pre-calculated
G
factor may be used when all other experimental parameters are
constant. In other cases, the system can measure the
G
factor automatically before an ex-
perimental run.
Constant Wavelength Analysis
Use
Constant Wavelength Analysis
to determine polarization at particular excita-
tion/emission wavelength-pairs.
Use Polarization Modes
checkbox ac-
tivates acquisition parameters for polariz-
ers.
This brings up the
Polarizer Settings
dialog box for polarization, anisotropy,
and raw polarization modes.
Proceed to Acquisitions
opens the
New
Polarization Sample
window: