Fluorolog-3 v. 2.2 (10 Sep 2002)
Components & Accessories
6-19
Warning:
Failure to clamp these hoses securely
may result in flooding and damage to the optics
and electronics of the instrument.
FL-1011 Four-Position Thermostatted Cell
Holder
The FL-1011 Four-Position Thermostatted Cell Holder keeps a sample at a constant
temperature from –20°C to +80°C. The temperature is maintained by an ethylene-
glycol–water mixture pumped through from an external circulating temperature bath
(not included). The holder also includes a magnetic stirrer, for mixing turbid or viscous
samples. Also required is the FM-2003 Sample Compartment Accessory.
FL-1011 Four-Position Thermostatted Cell Holder.
Installation
1
Remove the compartment gap-bed.
2
Position the FL-1011 gap-bed drawer.
3
Tighten with four screws.
4
Attach the ¼" tubing to the brass inlets on the
bottom of
the holder.