Fluorolog-3 v. 2.2 (11 Jul 2002)
System Maintenance
5-5
Note:
Jobin Yvon Inc. provides new bulbs with leads and
connectors of the proper length and size. Other manufac-
turers’ lamps may require extra adjustment before installa-
tion.
To insert a xenon lamp
1
Be sure the cover of the lamp housing is
removed.
2
Pay attention to the polarity of the xenon lamp.
Diagram of lamp emphasizing polarity.
3
Seat the xenon lamp in the bottom bracket.
Make sure that the lamp is inserted all the way into the aluminum bracket. The
negative end (marked with a “–“) should be downward.
4
Place the square aluminum upper bracket over
the positive terminal of the bulb.
Lamp lead
Positive end
Negative end
Lamp lead
Large electrode
Small electrode