Fluorolog-3 v. 2.2 (10 Sep 2002)
Components & Accessories
6-24
Note:
In the T-box sample com-
partment, the Model 1692M Se-
lection Mirror is an option. If
front-face measurements are to
be acquired, this mirror must be
installed in the sample com-
partment.
Note:
Before scanning a solid sample, Jobin Yvon Inc. recom-
mends running a water Raman scan in a cuvette with front-face
detection. This ensures accurate alignment of the Model 1692M
Selection Mirror.
For samples such as thin films, microscope slides,
fibers, or other materials:
1
Place the material on the block on the side
opposite that of the well.
2
Insert the block between the bracket and spring
clip.
The sample should be perpendicular to the excitation light and fluorescence
collected using front-face detection.
3
Select front-face
detection by
turning the knob
on the top panel
of the sample
compartment to
FF (front-face).