Fluorolog-3 v. 2.2 (10 Sep 2002)
Components & Accessories
6-21
Warning:
Failure to clamp these hoses securely
may result in flooding and damage to the optics
and electronics of the instrument.
FL-1012 Dual-Position Thermostatted Cell
Holder
The FL-1012 Dual-Position Thermostatted Cell Holder keeps a sample at a constant
temperature from –20°C to +80°C. The temperature is maintained by an ethylene-
glycol–water mixture pumped through from an external circulating temperature bath
(not included). The holder also includes a magnetic stirrer, enabling mixing of turbid or
viscous samples. Also required is the FM-2003 Sample Compartment Accessory.
FL-1012 Dual-Position Thermostatted Cell Holder.
Installation
1
Remove the present holder from the posts.
2
Replace with the FL-1012.
3
Tighten the two thumbscrews
4
Attach the ¼" tubing to the brass inlets on the
bottom of
the
holder.