Fluorolog-3 v. 2.2 (31 Jul 2002)
Producing Correction Factors
11-4
Warning:
Do not touch the
lamp. Use cotton gloves or
lens paper.
6
Attach the 2 wire ends of the standard lamp to
the lamp holder
posts.
a
The positive (+) lead goes to
the positive (red) side of the
holder.
b
The negative (–)
lead goes to the
negative (black)
side of the holder.
c
The nipple of the
lamp should point
upward.
7
Fix the lamp holder to the top of the sample
compartment with double-sided tape.
Make sure the the filament is vertically centered over the fixture.
View looking downward from the standard lamp to the mask. Dotted lines
(in perspective) indicate how the lamp should be over the square hole.
Nipple points upward