Fluorolog-3 v. 2.2 (11 Jul 2002)
System Operation
3-18
11
Close
Visual Instrument Setup
.
12
Confirm that the emission monochromator is
calibrated, by running another water Raman
scan in
Run Experiment
.
This time the peak should occur at 397 ± 0.5 nm.