Fluorolog-3 v. 2.2 (11 Jul 2002)
System Maintenance
5-10
Note:
If there is a problem, you may have to adjust the horizon-
tal position of the lamp. This is rarely necessary, and should
only be performed as a last resort.
Horizontal
adjustment
Focus
Vertical
adjustment
Note:
To reach the
set screw, you may
have to fish around
with the Allen
wrench.
8
Insert a 5/64" Allen wrench inside the middle
port on the top of the lamp housing.
The Allen wrench is ready to adjust the vertical position of the lamp.
9
View the trace and make vertical adjustments by
rotating the wrench either left or right to increase
the signal.
10
Return to the
Run Experiment
application.
11
Re-scan the water Raman spectrum.
This should result in a clean scan at maximum intensity.