Fluorolog-3 v. 2.2 (11 Jul 2002)
System Maintenance
5-9
Note:
If the lamp is not at
the best vertical position,
correctly positioning the
lamp will increase the sig-
nal significantly.
3
Enter the
Real Time Display
application.
4
Set the excitation and emission
spectrometers to 350 nm and
397 nm, respectively.
The Raman band’s emission appears at 397 nm
when water is excited using 350 nm.
5
Make sure HV1 is on and set to
950 V.
6
Open the programmable
excitation shutter.
7
View the incoming data in acquisition mode S
(signal detector).
Watch the signal intensity as you
adjust the vertical position of the
lamp. Therefore, position the monitor
to see the changes while standing by
the lamp housing.
Is the water Ra-
man peak at 397
± 0.5 nm?
Calibrate the emis-
sion monochroma-
tor.
Go to System
Operation.
No
Yes
Is the peak inten-
sity as listed in the
above table?
Yes
No
Lamp is
properly
adjusted.
Continue
with step
3 below.
Start