Fluorolog-3 v. 2.2 (11 Jul 2002)
Reassembly Instructions
13-2
Note:
See Chapter 2,
Sys-
tem Description
, for con-
figurations. Concerning
the lamp housing, xenon
sources, and cautions,
see Chapter 5,
System
Maintenance
.
Spectrofluorometer assembly
Once the computer has been assembled, the external components of the Fluorolog
®
-3
spectrofluorometer must be connected. The modules of the Fluorolog
®
-3 fit together in
a seamless configuration. Each module and the lamp housing has alignment studs,
alignment receptacles, and D-connectors on the outside of the housing:
1
Assemble the spectrofluorometer system’s
modules into the proper configuration.
That is, attach the spectrometers to the sample compartment, and the lamp
housing to the excitation monochromator.
Example of a Fluorolog
®
-3 configuration.
Source
Excitation
spectrometer
Sample compartment
module
Emission
spectrometer
Detector
Stud receptacle
Stud
Male D-connector
Female D-connector