Fluorolog-3 v. 2.2 (11 Jul 2002)
Reassembly Instructions
13-6
Note:
With special options (e.g.,
trigger accessory), there may be
an additional cable connection.
Refer to the accessory’s instruc-
tions.
4
Take the end of the #33977 cable that has two
connectors.
5
Plug the end with the 4-pin circular connector
into the POWER jack of the DM302, and the
BNC plug into the OUT jack of the DM302.
6
Plug the remaining end of the #33977 cable into
the INPUT S connector on the back panel of the
SAC.
Sample compartment module
1
Insert the 25-pin–9-pin cable
#400108 between the
sample compartment and
the jack labeled “mono
drive” on the SAC.
Lamp power supply
The lamp power supply is integrated into the lamp housing. No additional connections
are required.
Computer to SAC
The spectrofluorometer system is controlled via the SAC through the computer.
1
Connect the 9-pin–9-pin connector cable
between COM1 of the computer and COM1 of
the SAC.