Fluorolog-3 v. 2.2 (9 Aug 2002)
TRIAX operation with the Fluorolog
®
-3
14-32
Note:
The
Counts
window does not appear in
Real Time Display
with a
CCD detector, because the system is collecting data over many pix-
els, not reporting a single number.
Note:
Some slits can-
not be set in
Real
Time Display
.
Note:
The exit slit-
width parameter is
valid only for PMT
users. For CCD de-
tectors, there is no slit
at the exit of the
TRIAX, so any dis-
played settings are
superfluous.
The
Toolbar
When a single-point detector (e.g., PMT, InGaAs, IR, etc.) is attached to a TRIAX and
activated in the correct layout, an
HV
button appears in the
Toolbar
.
The
Intensity
window
The
Intensity
window is used
only
for a single-point detector
(e.g., PMT, InGaAs, IR, etc.).
The
Slits
window
Note the following terms used in
the
Slits
window:
•
Front
= axial slit
•
Side
= lateral slit