Fluorolog-3 v. 2.2 (9 Aug 2002)
TRIAX operation with the Fluorolog
®
-3
14-31
Real Time Display
Unlike typical systems,
Real Time Display
with a TRIAX and CCD must be used with
the
Run Experiment
window open simultaneously. The
Real Time Display
is used to
optimize the CCD detector and TRIAX settings for the best experimental results.
To do a CCD acquisition in the
Real Time Display,
1
In
Instrument
Control Center
,
click the Run Ex-
periment button.
The DataMax main window appears.
2
Choose Collect.
3
A drop-down menu appears.
4
Choose RTD-CCD.
The
Real Time Display
opens.