Fluorolog-3 v. 2.2 (9 Aug 2002)
TRIAX operation with the Fluorolog
®
-3
14-29
g
Choose the
Position
in-
dex-card tab.
The spectrum must
contain five known
peaks on different ar-
eas of the CCD chip.
h
Set
Nb
Region
to 1.
i
Click
OK
.
The
Acquisition
Setup
window closes.
In the
CCD
Acquisition
dialog
box,
j
Enter a file name.
k
Enter the integration time.
l
Enter other required parameters, e.g., slit widths.
m
Click
Run
.
If defined correctly for either technique, DataMax collects the spectrum. The
Select
Calibration Peak
window appears.
A dialog box appears requesting the
calibration wavelengths (for
Multi
Scans
) or verification (for
One
Scan
). When enough peaks have
been selected for calibration, the
Calc
button appears:
5
Select Calc
if correct;
select
Correct to adjust the values.
DataMax calculates new parameters.
6
Close and exit all DataMax programs.
7
Restart DataMax to cause the changes to take
effect.