TRIAX operation with the Fluorolog
®
-3
14-24
Flush
The
Flush
option can remove all residual charges from the CCD chip before data ac-
quisition. The drop-down menu contains three choices:
•
Flush Before Each
Acquisition
Clear out all charges before an accumulation.
•
No Flush
Charges are not removed before an accumu-
lation. This is faster, but background noise
may increase between acquisitions. For mul-
tiple scans, the first acquisition may have a
higher background-noise level than later
ones.
•
Flush Before First
Scan
The charges are removed only before the first
scan.