Fluorolog-3 v. 2.2 (11 Jul 2002)
System Description
2-10
Fluorolog®-3 Model FL3-XXX
Systems with the T-configuration design were developed for T-format polarization or
anisotropy and dual-emission spectroscopy. Models FL3-XXX (where the Xs are the type of
spectrometer positioned at the excitation, first-emission, and second-emission positions,
respectively) are available in numerous configurations.
Excitation
Spectrometer
Source
Sample
Compartment
Module
Emission
Spectrometer
Emission
Spectrometer
Dectector
Dectector
Fluorolog
®
-3 model FL3-122
The T-configuration allows the Xe source and the signal detector to be positioned at right-
angles to the sample, for a variety of experiments with results unequaled by other
configurations. The optical configuration of the model FL3-122 is shown below.
PMT
PMT
Source
Single-grating
Excitation Spectrometer
Double-grating
Emission Spectrometer
Double-grating
Emission Spectrometer
T-box Sample
Compartment Module
Optical layout of the Fluorolog
®
-3 Model FL3-122.
Detector
Detector