4.9 Setting Pre-Measurement Delay
86
Setting the Delay Time
Set the delay so that reactance component (inductance or capacitance) does not affect
measurements.
To fine tune the delay, begin with a longer delay than necessary, then gradually shorten it
while watching the measured value.
1
Open the Settings Screen.
2
Open the Measurement Setting Screen.
3
Select whether to use the preset (default) or a user-set value.
The Settings screen
appears.
Switch the function menu
to P.2/3.
1
2
Move the cursor to the
[MEAS] tab with the left and
right cursor keys.
2
Preset (internal fixed value)
(go to step 5)
User-set
Selection
1