3.5 Checking Measured Values
59
3
Setting the measurement method for an open target (current fault mode
setting)
This section describes how to configure instrument operation when current fault output is
detected.
When set to current fault, a break in the measurement target wiring is determined to be an
error, and no comparator judgment is made. When set to over-range, a break in the mea-
surement lead or other open state is determined to be an over-range event, and a compar-
ator judgment of Hi results. Choose the setting that best suits your application.
The current fault mode setting applies to all channels. (RM3545-02 only)
1
Open the Settings Screen.
2
Open the Measurement Setting Screen.
3
Select the desired current fault mode.
4
Return to the Measurement screen.
The Settings screen
appears.
Switch the function menu
to P.2/3.
1
2
Move the cursor to the
[MEAS] tab with the left and
right cursor keys.
2
Current fault (default)
Over-range
Selection
1
Return to the
Measurement screen.