4.12 Maintaining Measurement Precision (Self-Calibration)
94
Auto setting operation
Self-calibration starts immediately after measurement completes and is finished in 5 ms.
One TRIG signal received during self-calibration is held, and measurement will start after
the self-calibration completes.
Additionally, self-calibration is performed once every second during TRIG standby opera-
tion. In the event the TRIG signal is received during self-calibration, the self-calibration is
canceled, and measurement will start after 0.5 ms.
•
During auto-scan operation, self-calibration starts only after scanning completes. Self-cal-
ibration will not be performed after each channel is measured.
•
A 400 ms self-calibration is performed immediately after switching from MANUAL to
AUTO. Do not input the TRIG signal during that interval.
If there is at least 5 ms of extra time in the measurement interval
If the TRIG signal is received during self-calibration
Measurement
Self-
Calibration
Self-
Calibration
Self-
Calibration
Self-
Calibration
The TRIG signal received during
self-calibration is held.
Measurement
processing
Measurement
processing
Measurement
Measurement
Measurement
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......