
8.3 Multiplexer Settings
149
8
3
Set the measurement terminals.
4
Select the measurement method.
When the measurement method is switched, the multiplexer channel settings will be
initialized (i.e., a multiplexer channel reset will be performed). Always finalize the
measurement method before allocating pins or performing zero-adjustment.
5
Set the scan function.
When the scan function is set to auto or step, external trigger operation will be used
regardless of the trigger source setting.
2
Make measurements using
the front measurement ter-
minals (do not use the multi-
plexer). (default)
Use the multiplexer.
Selection
1
2
4-wire (default)
2-wire
Selection
1
2
Use auto-scan
(measure all channels at
each TRIG signal).(default)
Use step scan
(measure 1 channel at each
TRIG signal).
Do not scan.
Selection
1
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......