Chapter 13 Specifications
266
(19)BIN Function
* When set using the instrument’s keys, the input range will reflect the range and scaling coefficient with a max-
imum resolution of 1 n
Ω
and a maximum value of 9 G
Ω
.
(20)Comparator Beeper Setting
(21)Auto Hold Function
Operation
Compares setting and measured values and displays the result.
Setting
ON/ OFF (When the BIN function is on, the range and comparator functions are
fixed to off. When
Δ
T is on while using the MUX measurement terminal setting,
the BIN function is automatically turned off.)
Comparator mode
ABS mode/ REF% mode
Display
Absolute value (resistance value) display only
BIN number
0 to 9
Default state
OFF
Judgment
Hi
Measured value > Upper limit value
IN
Upper limit value
≥
measured value
≥
Lower limit value
Lo
Lower limit value > measured value
ABS Mode
Upper/Lower limit
ranges
0.000 0 m
Ω
to 9000.00 M
Ω
*
Default setting
0.000 0 m
Ω
REF% Mode
Reference value range
0.000 1 m
Ω
to 9000.00 M
Ω
*
Upper/ Lower limit
ranges
0.000% to ±99.999%
Default setting
Reference value: 0.000 1 m
Ω
, Upper/ Lower limit ranges: 0.000%
Operation
Sounds a beeper based on the comparator judgment result or total judgment.
(Set separately for Hi/ IN/ Lo and for PASS/ FAIL when using the MUX mea-
surement terminals.)
Operation settings and
tones
type 1/ type 2/ type 3/ OFF
Number of beeps
1 to 5 times / continuous
Default setting
OFF, 2times
Operation
Holds measured values automatically (only when using the measurement ter-
minals on the front of the instrument with the internal trigger source and con-
tinuous measurement on [free-run]).
The hold is canceled when the measurement leads are removed from the tar-
get and the next measurement performed, or when the
key is pressed.
Operation setting
ON/ OFF
Default setting
OFF
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......