10.2 Timing Chart
190
•
For pulse EOM output, the EOM signal turns OFF when the pulse width time elapses.
•
When using the internal trigger [INT] setting, the EOM signal consists of pulse output with
a width of 5 ms. The ERR signals do not turn OFF at the start of measurement. They are
updated at the completion of the next measurement.
•
When not using the multiplexer, the zero-adjustment time is approximately 600 ms when
using a manually set range and approximately 4 s when using auto-ranging. When per-
forming scanning zero-adjustment while using the multiplexer, the zero-adjustment time
will elapse for each channel.
Zero-adjustment timing
ON
0ADJ
Zero-adjustment in progress
OFF
Zero-adjustment
processing
INDEX
EOM
ERR
OFF
OFF
OFF
ON
ON
10 ms
or more