Chapter 13 Specifications
260
(10)Delay Setting
Operation
Adjusts the time for measurement to stabilize by inserting a waiting period after
using the OVC or the auto-range function to change the measurement current
or after the TRIG signal.
Preset: Integration starts after the factory-default time (which varies with the
range) elapses.
User-set: Integration starts after the specified time elapses (for all ranges).
Setting
Preset (internal fixed value)/ user-set (set value)
Delay setting range
0 ms to 9999 ms
Default setting
Preset/ 0 ms
Preset delay value (internal fixed) (unit: ms)
LP OFF
LP ON
Range
100 M
Ω
range
high-precision
mode
Measure-
ment cur-
rent
Delay
OVC: OFF
OVC: ON
10 m
Ω
−
−
75
25
100 m
Ω
−
High
250
25
−
Low
20
2
1000 m
Ω
−
High
50
2
−
Low
5
2
10
Ω
−
High
20
2
−
Low
5
2
100
Ω
−
High
170
2
−
Low
20
2
1000
Ω
−
−
170
2
10 k
Ω
−
−
180
−
100 k
Ω
−
−
95
−
1000 k
Ω
−
−
10
−
10 M
Ω
−
−
1
−
100 M
Ω
ON
−
500
−
100 M
Ω
OFF
−
1
−
1000 M
Ω
OFF
−
1
−
Delay
1
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......