14.1 Troubleshooting
285
14
2. Measurement issues
1-4 The LED Com-
parator Attach-
ment will not turn
on.
Instrument’s com-
parator lamp
On
The attachment is not properly con-
nected.
→
Connect the LED Comparator
Attachment properly to the
COMP.OUT jack.
There is a broken connection.
→
Replace the LED Comparator
Attachment.
−
Off
See No. 1-3 above, "The instrument’s
comparator lamp will not turn on."
1-5 The beeper is not
audible.
Key operation
sound setting
OFF
The function is OFF.
→
Turn ON the function.
Judgment sound
setting
OFF
The function is OFF.
→
Turn ON the function.
1-6 You wish to
change the
beeper volume.
The instrument’s beeper volume
cannot be changed.
−
−
No
Issue
Items to check
Possible causes
→
Solutions
See
2-1 Measured val-
ues are unstable.
Effects of noise
Susceptibility to
noise
Measurement
leads
Clip-type leads
See Appendix 7(3).
Wiring becomes
two-terminal wir-
ing in middle.
Measurement tar-
get
Wide or thick
See Appendix 7(4).
Temperature is
unstable (just
manufactured,
just opened,
being held by
hand, etc.).
Low heat capac-
ity
Transformer,
motor, choke
coil, solenoid
No
Issue
Items to check
Possible causes
→
Solutions
See
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......