14.1 Troubleshooting
292
5. Printer issues
6. Multiplexer issues
No
Issue
Items to check
Possible causes
→
Solutions
See
5-1 No data is being
printed.
The printer is not connected.
→
Check whether the connector
has been connected.
→
Check whether the interface set-
ting is correct.
If using the PRINT signal, see No.
3-3 above.
5-2 Printed text is
garbled
The printer and instrument set-
tings do not match.
→
Check the printer settings again.
No
Issue
Items to check
Possible causes
→
Solutions
See
6-1 It is not possible
to switch to the
multiplexer
inputs.
Display
ERR:60
Measurement leads are connected
to the measurement terminals on
the front of the instrument.
→
Do not connect any measure-
ment leads to the measurement
terminals on the front of the
instrument.
If
ERR:60
is displayed even
though no measurement leads
are connected, turn off the
instrument and remove the
Z3003. If
ERR:60
is no longer
displayed after removing the
Z3003, the Z3003 may be bro-
ken. Have it repaired.
6-2 Channels cannot
be switched by
operating the
instrument’s
keys.
CH
is not being
displayed.
The front terminals are being used
as the measurement terminals.
→
Set the measurement terminals
to MUX.
Scan display (set
to list display)
The scan function is set to auto or
step.
→
Set the scan function to OFF in
order to switch channels with
key operation.
The set unit number and the unit
number in which the Z3003 is
installed differ.
→
Check the settings and the unit
on the back of the instrument.
RMT
is being dis-
played.
The instrument is in remote mode,
in which it is controlled by commu-
nications functionality.
→
Operate the instrument after
canceling remote mode.
6-3 Channels cannot
be switched with
EXT I/O.
The MUX signal is not on.
→
Turn on the MUX signal.
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......