5.4 Performing Temperature Rise Test (Temperature Conversion Function (
Δ
T))
117
5
1
Open the Settings Screen.
2
Open the Measurement Setting Screen.
3
Enable the temperature conversion function. (
Δ
T)
4
Set the initial resistance and initial temperature.
Set the initial resistance and initial temperature in Steps
1
through
3
.
The Settings screen
appears.
Switch the function menu
to P.2/3.
1
2
Move the cursor to the
[MEAS] tab with the left and
right cursor keys.
Enables the function
Disables the function
(default) (go to step 6)
Selection
2
1
Setting range initial resistance
: 0.001
μΩ
to 9000.000 M
Ω
(default: 1.0000
Ω
)
initial temperature : -10.0 to 99.9°C (default: 23.0°C)
Move the cursor to the setting you
wish to configure. Make the value
editable with the
key.
2
Move the cursor to the digit you
wish to set with the left and right
cursor keys. Change the value
with the up and down cursor keys.
Change
values.
1
Move among
digits.
3
Accept
( Cancel)
The initial resistance value range varies with the scaling setting.
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......