11.4 Auto-Exporting Measured Values (at End of Measurement) (Data Output Function)
235
11
Output data format
Measured value format when scaling is off
(The measured value format varies depending on scaling. (p.77))
Changing the number of digits in the measured value will not change the format. Undis-
played digits have a value of 0.
For positive measured values, a space (ASCII 20H) represents the “+” sign.
When ±OvrRng is displayed, values are ±1E+20.
When a measured value fault occurs, values are ±1E+30.
•
This function is not applicable to the GP-IB Interface.
•
This function cannot be used when the scan function is set to auto or step while using the
MUX measurement terminals.
•
When using the internal trigger [INT], data is automatically sent at TRIG signal input or
when the
key is pressed.
•
Do not use commands when data output is ON. Doing so may cause measured values to
be sent twice or other issues.
•
Resistance value (Absolute value display, unit:
Ω
)
Low-
power
Measurement
Range
Measured Value
±OvrRng
Measurement Fault
OFF
10 m
Ω
±
.
E-03
±10.19
+10.28
100 m
Ω
±
.
E-03
±100.0000E+18
+100.0000E+28
1000 m
Ω
±
.
E-03
±1000.000E+17
+1000.000E+27
10
Ω
±
.
E+00
±10.19
+10.29
100
Ω
±
.
E+00
±100.0000E+18
+100.0000E+28
1000
Ω
±
.
E+00
±1000.000E+17
+1000.000E+27
10 k
Ω
±
.
E+03
±10.19
+10.29
100 k
Ω
±
.
E+03
±100.0000E+18
+100.0000E+28
1000 k
Ω
±
.
E+03
±1000.000E+17
+1000.000E+27
10 M
Ω
±
.
E+06
±10.19
+10.29
100 M
Ω
±
.
E+06
±100.0000E+18
+100.0000E+28
1000 M
Ω
±
.
E+06
±1000.000E+17
+1000.000E+28
ON
1000 m
Ω
±
.
E-03
±1000.00E+17
+1000.00E+27
10
Ω
±
.
E+00
±10.0000E+19
+10.0000E+29
100
Ω
±
.
E+00
±100.000E+18
+100.000E+28
1000
Ω
±
.
E+00
±1000.00E+17
+1000.00E+27
•
Resistance value (Relative value display, unit: %)
Measured Value
±OvrRng
Measurement Fault
±
.
E+00
±100.000E+18
+100.000E+28
•
Temperature, temperature conversion display (unit: °C)
Measured Value
±OvrRng
Measurement Fault
±
.
E+00
±100.0E+18
+100.0E+28
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......