5.2 Classifying Measurement Results (BIN Measurement Function)
108
BIN Measurement compares a measured value with up to ten sets of upper and lower
thresholds (BIN 0 to BIN 9) in one operation, and display the results. Measured values that
do not fall in any BIN are judged to be OB (out-of-bin). Judgment results are output at the
EXT I/O connector.
See:
"Connector Type and Signal Pinouts" (p. 177)
•
When the BIN measurement function is on, the comparator cannot be turned on.
•
Turning on
Δ
T or setting the measurement terminal to multiplexer automatically turns off
the BIN measurement function.
•
The range cannot be changed while using the BIN measurement function. To change the
range, do so with the
and
keys on the BIN Number Settings screen.
Turn off the BIN measurement function when using auto-ranging.
5.2
Classifying Measurement Results
(BIN Measurement Function)
1
Open the Settings Screen.
2
Open the BIN Setting Screen.
3
Open the BIN Setting Screen.
The Settings screen
appears.
Switch the function menu
to P.2/3.
1
2
Move the cursor to the [BIN]
tab with the left and right cur-
sor keys.
Enables the BIN function
Disables the BIN function
(default)
Selection
2
1
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......