14.1 Troubleshooting
287
14
2-3 No measured
value is dis-
played.
(Concerning the
display of mea-
sured value
faults, see also
p.55.)
Measured values
- - - - - - -
There is a break in the measurement
leads.
→
Replace the measurement leads.
The contact resistance is too high (for
user-made leads).
→
Increase the contact pressure.
→
Clean or replace the probe tips.
→
Use a range with a low measure-
ment current or set the measure-
ment current to low.
The wiring resistance is too high (for
user-made leads).
→
Make the wiring thicker and shorter.
→
Use a range with a low measure-
ment current or set the measure-
ment current to low.
CONTACT
TERM.A
,
CONTACT
TERM.B
The probe is worn.
There is a break in the measurement
leads.
→
Replace the measurement leads.
The probe is not coming into contact
with the measurement target.
→
Place the probe in proper contact
with the target.
−
The resistance value between the
SENSE and SOURCE is high
because the measurement target is
conductive paint, conductive rubber,
or a similar material.
→
Turn the contact check function off.
OvrRng
The measurement range is low.
→
Select a high-resistance range or
use auto-ranging.
SW.ERR
ERR:061
A multiplexer relay hot-switching pre-
vention function error has occurred.
→
The relay cannot be switched
because the current from the mea-
surement target has not
decreased. Increase the delay set-
ting since the measurement circuit
may be being influenced by back
EMF from a transformer or other
device. Do not apply any current or
voltage to the measurement termi-
nals.
NO UNIT
No multiplexer unit has been inserted.
→
Insert the unit properly. Do not allo-
cate units that have not been
inserted to channels.
No
Issue
Items to check
Possible causes
→
Solutions
See
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......