3.3 Setting the Measurement Speed
50
The measurement speed can be set to FAST, MED (medium), SLOW1, or SLOW2. The
MED (medium), SLOW1, and SLOW2 settings offer increased measurement precision
compared to the FAST setting as well as greater resistance to the effects of the external
environment.
If the setup is excessively susceptible to the effects of the external environment, shield the
measurement target and measurement leads adequately and twist the cables together.
See:
"Appendix 8 Mitigating Noise" (p. A19)
A self-calibration that lasts about 5 ms is performed between measurements. To shorten
the measurement interval, set the self-calibration to “manual.”
See:
"4.12 Maintaining Measurement Precision (Self-Calibration)" (p.92)
Integration time (Unit: ms) (detected voltage data acquisition time)
Shortest measurement times when using the internal trigger source with
continuous measurement on (free-run)
3.3
Setting the Measurement Speed
Press this to change the measurement speed.
When OVC is on, integration is performed twice. When LP is on, OVC is fixed to on.
When using the SLOW2 measurement speed with low-power resistance measurement on, the in-
strument will performing averaging with two iterations internally even if the averaging function is
set to off.
* When using the MUX mesurement terminals, the integration time is 1.0 ms only in the 10 m
Ω
range.
See:
"13.1 Instrument Specifications" (p.249)
LP
Range
FAST
MEDIUM
SLOW1 SLOW2
50 Hz 60 Hz 50 Hz 60 Hz
OFF
1000 k
Ω
or less
0.3*
20.0
16.7
100
200
10 M
Ω
or more
20.0
16.7
20.0
16.7
100
200
ON
All ranges
20.0
16.7
40.0
33.3
200
300
Shortest conditions
Delay: 0 ms, TC: ON, Comparator: ON, OVC: OFF, Average: OFF,
Self-Calibration: MANUAL, Contact improver: OFF, Scaling: OFF
Measured value display switching: None
* When using the MUX mesurement terminals, the shortest measurement time is 1.7 ms only in
the 10 m
Ω
range.
LP OFF (unit: ms), tolerance: ±10%±0.2 ms
Range
FAST
MEDIUM
SLOW1 SLOW2
50 Hz 60 Hz 50 Hz 60 Hz
1000 k
Ω
or lower range
1.0*
20.7
17.4
101
201
10 M
Ω
or greater range
20.7
17.4
20.7
17.4
101
201
LP ON (unit: ms), tolerance: ±10%±0.2 ms, Only with OVC on
Range
FAST
MEDIUM
SLOW1 SLOW2
50 Hz 60 Hz 50 Hz 60 Hz
1000 m
Ω
71
65
111
98
431
631
10
Ω
111
105
151
138
471
671
100
Ω
111
105
151
138
471
671
1000
Ω
113
107
153
140
473
673
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......