10.2 Timing Chart
197
10
(2) Scan function: Auto
Measurement is performed while switching all channels after one trigger input.
•
The channel judgment result (HI, IN, LO, ERR) signals and BCD signal are not output.
Only the judgment result (T_PASS, T_FAIL, T_ERR) signals are output.
•
The INDEX signal does not turn on for each channel. It turns on after the completion of
scanning.
•
During scanning, the TRIG, CAL, and 0ADJ signals are ignored without being held.
SCN_STOP operation
State
Judgment result/
BCD
Switching time
Switching
processing
Channel 1
Measure-
ment
Switching
processing
Switching
processing
Channel 20
Measure-
ment
Switching
processing
Channel 1
Measure-
ment
Judgment result/ BCD: HI, IN, LO, ERR, PASS, FAIL, BCDm-n, RNG_OUT0 to 3
In this example, channels 1 through 20 have been set to ON.
State
10 ms or more
Approx. 15 ms
Switching
processing
Channel 1
Measure-
ment
Switching
processing
Channel 2
Measure-
ment
Switching
processing
Channel 3
Measure-
ment
Rest
processing
Switching
processing
Channel 1
Measure-
ment
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......