4.3 Zero Adjustment
69
4
•
If a resistance that is smaller than the resistance value when zero-adjustment was per-
formed is measured, the measured value will be negative.
Example: If you set an offset of 50 m
Ω
for the 100 m
Ω
range
→
If you measure 30 m
Ω
, -20 m
Ω
will be displayed.
•
When using the multiplexer, zero-adjustment can be performed by scanning all channels.
See:
"8.5 Zero Adjustment (When a Multiplexer Unit Has Been Installed)" (p.163)
Allow the instrument to warm up for 60 minutes before performing zero-adjustment.
Performing zero-adjustment
1
Short the measurement leads together.
Red
SOURCE
SOURCE
SOURCE
SENSE
SENSE
SENSE
Red
Black
Black
SOURCE
SENSE
Bring the "V" marks together
at the same position.
Connection
Connection
SENSE A SENSE B
SOURCE A
SOURCE B
L2104 (option)
L2101
L2102, L2103 (options)
Since zero-adjustment cannot be performed with the L2102 or L2103,
use the L2101 Clip Type Lead or other lead type to perform zero-ad-
justment.
Place the alligator clips on the outside
and the lead rods on the inside when
performing zero-adjustment.
Correct
Incorrect
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......