14.2 Replacing the Measurement Circuit’s Protective Fuse
300
Inserting the fuse holder without first placing a replacement fuse into it may make it difficult
to remove the fuse holder. Be sure to load a replacement fuse before inserting the holder.
14.2 Replacing the Measurement Circuit’s
Protective Fuse
•
Replace the fuse only with one of the specified type, characteristics,
rated current, and rated voltage.
Do not use fuses other than those specified (especially, do not use a
fuse with higher-rated current) or do not short circuit and use the fuse
holder. Doing so may damage the instrument and result in personal
injury.
Fuse type: F1.6AH/250V (non-arcing) 20 mm × 5 mm dia.
•
To avoid electric shock, turn off the main power switch and discon-
nect the cords and leads before replacing the fuse.
Rear panel
Fuse holder
1
Confirm that the instrument's Main
power switch (rear panel) is OFF(
),
and disconnect the power cord.
2
Unlock the fastener on the fuse holder
on the rear panel using a slotted screw-
driver, and remove the fuse holder.
3
Replace the fuse with a rated fuse. (The
replacement method may differ depend-
ing on the shape of the fuse holder.)
4
Reset the fuse holder.
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......