Chapter 13 Specifications
263
13
(15)Self-Calibration Function
(16)Contact Improvement Function
(17)Faulty Measurement Detection
Operation
Compensates for offset voltage and gain of measurement circuit
Setting
AUTO/ MANUAL
Compensation timing
AUTO
: At power-on, after measured value, during TRIG standby (every 1 s)
MANUAL: During EXT I/O CAL signal input, when executing the calibration
command
Self-calibration time
At power-on, when switching to auto and during manual execution: 400 ms
Auto: 5 ms (moving average)
Default setting
AUTO
Operation
A voltage is applied between the SENSE A and SENSE B terminals after TRIG
signal input, and a contact improvement current is allowed to flow for 0.2 ms.
Setting
OFF/ ON (When LP is on, the contact improvement function is fixed to off.)
Default setting
OFF
Applied voltage
Max. 5 V
Contact improvement
current
Max. 10 mA (flowing to the measurement target)
Over Detection Function
Operation
Indicates under- or over-range values in the following conditions:
• Measured value is outside of the measurement range
• Measured value is outside of the A/D converter input range
• Calculation result exceeded the number of display digits
Contact Check Function
Operation
Checks the connections between SOURCE A and SENSE A, and between
SOURCE B and SENSE B terminals
Setting
ON/ OFF (When using the MUX measurement terminal setting with the 2-wire
measurement method, fixed to off. When using the 100 M
Ω
or greater range,
the setting is fixed to ON.)
Threshold
50
Ω
(reference value)
Default setting
ON (When LP is off), OFF (When LP is on)
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......