![Hioki DM7275 Series Скачать руководство пользователя страница 73](http://html1.mh-extra.com/html/hioki/dm7275-series/dm7275-series_instruction-manual_2137275073.webp)
70
Auto-hold Function
6.2
Auto-hold Function
The auto-hold function is useful when you wish to check the measured value. The buzzer will sound
once the measured value has stabilized (when the fluctuations in the measured value fall within the
auto-hold range), and the display will be automatically held. The auto-hold range is specified as a
percentage of the measurement range. Increasing the auto-hold range causes values to be held
more quickly, while decreasing it takes more time but causes values to be held in a more stable
state.
(Measurement screen)
MENU
>
MEAS
>
AUTO HOLD
1
Enable the auto-hold function.
(Default setting:
OFF
)
2
1
2
3
Set the auto-hold range.
CLR
Reverts the setting to its
default value.
CANCEL
Cancels the setting and
returns to the previous screen.
(Valid setting range: 0.001% to 1.000% of
the range; default setting: 0.1%)
The
HOLD
icon will be displayed on the
Measurement screen while the measured
value is being held automatically.
The current measured value is displayed in real
time.
Measured value being automatically held
•When the auto-hold function is enabled, the
measurement conditions change as follows:
RUN
state,
MEDIUM
integration time,
10 M
Ω
input resistance, contact check
ON
.
•Measured values are not automatically held
when they are 0.1% or less of the range.
When measuring small values, select an
appropriate range.
The auto-hold state will be canceled in
the following instances:
• If the measurement cables are
disconnected from the measurement
target and then reconnected to the
measurement target
• If the range or measurement speed
(integration time) is changed