176
Error Displays
14.3 Error Displays
The following messages are displayed on the screen when the instrument malfunctions or
encounters an abnormal measurement state.
• If you feel that the instrument may be malfunctioning, contact your authorized Hioki distributor or
reseller after reviewing the information provided in “Q&A (Frequently Asked Questions)” (p. 168).
• When an error is displayed on the LCD screen and service is required, please contact your
authorized Hioki distributor or reseller.
Display
Description
Solution
+OvrRng/−OvrRng
Over-range
Select the appropriate range.
NoCntct
Contact error
Check the connections with
the object to be measured.
Or, adjust the threshold for the
Err.TC
Temperature compensation
error
Connect a temperature sensor.
ERR:001
Lower limit is higher than
Upper limit.
Cannot set because the lower
limit value is greater than the
upper limit value.
Set an upper limit value that
is greater than the lower limit
value.
ERR:004
Unable to change the
setting during auto-hold.
When auto-hold function
is enabled, settings for
measurement speed and
continuous mode cannot be
changed.
Turn OFF the auto-hold
function. (p. 70)
ERR:005
Unable to set NULL
due to an abnormal
measurement value.
When
OverRng
,
NoCntact
, or
------
is displayed, NULL value
cannot be obtained.
Return from the abnormal
ERR:030
Command error.
Remote command syntax
error. (String is incorrect or
incorrect character code is
used.)
Check if the commands are
correct. (See the application
disc provided.)
ERR:031
Execution error. Invalid
parameter.
Remote command execution
error. The parameter value is
out of range.
Check if the parameters are
correct.
ERR:032
Execution error.
Remote command execution
error.
Check the execution error
conditions for each command.
ERR:050
The panel does not exist.
Panels that have not been
saved cannot be read.
Select a proper panel. (p. 61)
ERR:051
The panel does not exist.
Unable to rename.
Name of panels that have
not been saved cannot be
changed.
Select a panel that has been
saved. (p. 61)
ERR:060
Cannot use USB memory.
Set I/F function to USB-
MEMORY.
When
I/F
is set to
USB COM
,
USB flash drive cannot be
used.
Set the I/F to
USB MEMORY
.
ERR:061
The drive is not ready.
(No USB memory
inserted)
USB flash drive is not inserted.
Insert a USB flash drive.
ERR:062
This format is not
supported.
Format of the USB flash drive
is not correct.
Format the USB flash drive to
FAT32.
www.
.com
1.800.561.8187
Содержание DM7275 Series