![Hioki DM7275 Series Скачать руководство пользователя страница 128](http://html1.mh-extra.com/html/hioki/dm7275-series/dm7275-series_instruction-manual_2137275128.webp)
127
Connections (Instrument and Control Device)
11.3 Connections (Instrument and Control Device)
Before connecting, read “Before connecting to the EXT I/O connector” (p. 12) carefully.
The EXT I/O interface can be used to perform the following types of control:
Capability
Operations (signals)
(1)
Acquiring comparator judgment
results
Start measurement (TRIG signal)
↓
Measurement complete (EOM signal)
↓
Acquire judgment results (HI, IN, LO, ERR signals)
(2)
Acquiring BIN judgment results
Start measurement (TRIG signal)
↓
Measurement complete (EOM signal)
↓
Acquire measured values (BIN0 to BIN9 signals, OB signal,
ERR signal)
(3)
Loading panel data
Specify panel (LOAD0 to LOAD4 signals)
↓
Start measurement after panel load operation (TRIG signal)
(4) General-purpose input and output
:IO:INPut?
command (IN0, IN1 signals)
:IO:OUTPut?
command (OUT0 to OUT7 signals)
(5) Key lock
Enable key lock (KEY_LOCK signal)
(6)
Printing
Print (PRINT signal)