![Hioki DM7275 Series Скачать руководство пользователя страница 169](http://html1.mh-extra.com/html/hioki/dm7275-series/dm7275-series_instruction-manual_2137275169.webp)
170
Q&A (Frequently Asked Questions)
No.
Issue
Items to check
Possible causes → Solutions
See
2-2
Measured value
differ from
expected value. (A
negative value is
displayed.)
Scaling
function
ON
The offset setting is not correct.
→ Turn scaling
OFF
, or reconfigure
the setting properly.
Measurement cable connection
Cable is not connected properly.
→ Check the connections.
See also No. 2-1.
NULL function
ON
Zero point is shifted.
→ Set the NULL function
OFF
, or
reconfigure the settings properly.
2-3
Measured value is
not displayed.
(
For more
information about
measurement
error displays, see
Measured
value
NoCntct
There is a break in the
measurement cable.
→ Replace the measurement
cables.
Metal pins (probes) of the
measurement cable are worn or
cables are cut.
→ Replace the measurement
cables.
The metal pins (probes) are not
in contact with the measurement
target.
→ Clean or replace the metal pins
(probes).
→ Increase the contact pressure.
The measurement target is made of
a material such as conductive paint
or conductive rubber, resulting in a
high resistance value between the
HIGH and LOW terminals.
→ Set the contact function to
OFF
,
or use a smaller threshold.
(When measuring enclosure
potential)
Capacitance between battery
electrode and enclosure is small.
→ Set the contact function to
OFF
,
or use a smaller threshold.
+OvrRng
−OvrRng
Measurement range does not cover
the object to be measured.
→ Change the range or set it to
auto-range.
Nothing is displayed. No range is selected during auto-
range operation.
→ See No. 2-4.
-
2-4
No range is
selected during
auto-range
operation (no
appropriate range
is found).
Circuit to be measured
Voltage is fluctuating.
→ Use a fixed range.
2-5
The auto-hold
function is not
working (hold
operation is not
being canceled).
Measured
value
Not stabilizing.
→ See No. 2-1.
Does not change.
Incorrect range.
→ Select an appropriate range or
use auto-range.