Data Output Settings
9.3 Data Output Settings
(Measurement screen)
MENU
>
I/F
1
Enable automatic output.
(Default setting:
OFF
)
When automatic output is set to
ON
, do not
use communications commands. Doing so
may cause measured value data to be sent
twice.
2
(To change the information that will be
output)
1
2
MEASUREMENT
ITEM
V
: Voltage value (default
setting)
V, TEMP
: Voltage value
and temperature
DATE-TIME
Measurement time and
date (default setting:
OFF
[output disabled])
COMP
CONDITION
ALL
: All judgments (default
setting)
HI
: HI judgments
IN
: IN judgments
LO
: LO judgments
HL
: HI and LO judgments
•If comparator judgment or BIN measurement
is set to
ON
, judgment results will also be
output.
•When the interface is set to
USB KEYBD
,
DATE-TIME
is not output.
3
(To change the output format)
See “7.8 Setting Output Formats” (p. 92).
9
Outputting Data