20
Operating the Instrument
Starting measurement
By default, the instrument is in the
RUN
state. In this state, measurement will continue
automatically.
Continuous measurement (default setting:
RUN
state)
See “Continuous measurement”(p. 37).
RUN
state
Measurement will continue
automatically, and measurement
data will be saved in the instrument’s
internal memory.
STOP
state
Measurement will stop, and
the last measured value will be
saved.
Making measurements at the desired timing
See “Trigger measurement (measurement with user-specified timing)”(p. 38).
Starting measurement
Measurement can be started by either of the
following methods:
•While in the
STOP
state,
•While the trigger source is set to
EXTERNAL
,
send the TRIG signal to the instrument from
an external device.
After the set number of measurements (default
setting: 1) have been performed, measurement
will stop automatically.
Measurement data will be saved in the
instrument’s internal memory.
Up to 5000 measured values can be saved in the instrument’s internal memory. Saved
measured values can be displayed in graph form to illustrate the trend in voltage readings
(trend display) or output to a USB flash drive.