31
3
Measurement
3.1 Connecting the Measurement Cables
(Measurement Target)
Before connecting the measurement cables to the measurement target, read “Before connecting a
measurement cable” (p. 9) and “Before starting a measurement” (p. 10) carefully.
Use Hioki’s optional test leads, contact pins, alligator clips, or other connectors as appropriate
depending on the measurement target.
For more information: “Options (Sold Separately)” (p. 3)
Using the L9207-10 Test Lead
(1)
About the L9207-10
4
Black
Red
5
1
3
2
The test lead plugs are fitted
with protective caps. Remove
the caps before use.
Protective cap
1
Metal pins
Connect the metal pins to the
measurement target.
With cap: 4 mm or less
Without cap: 19 mm or less
Thickness: Approx. 2 mm
2
Caps
Fit the caps to the metal pins to prevent
short-circuits. The test leads can also be
used with the caps removed.
3
Barriers
The barriers indicate the safe distance
from the metal pins.
During measurement, do not touch
the area in front of the barriers.
4
Plugs
Connect the plugs to the instrument’s
measurement terminals.
5
Cables
The cables have a double-insulated
design.
(Length: Approx. 900 mm; thickness:
approx. 3.6 mm)
If the white portion of the inside of
the cable is exposed, replace the
test leads with a new L9207-10 set.
Remove the caps when using the test leads with the L4933 Contact Pin Set or the L4934 Small
Alligator Clip Set.
Removing the caps
Fitting the caps
Gripping the base of the cap gently, pull
to remove.
Store caps after removal so as not to lose
them.
Pass the test lead metal pin through the
hole in the cap and insert firmly all the
way.
3 Measurement
3
Measurement
www.
.com
1.800.561.8187
Содержание DM7275 Series
Страница 207: ...13 09 www com information itm com 1 800 561 8187...