114
Data Output Settings
Output data format
Example: When the scaling function is OFF, the number of display digits is 7.5, and the output
format is set to the decimal period
(The output data format varies depending on the scaling function setting, number of display digits
setting, and output format setting.)
See “Correcting measured values using a linear expression (scaling function)” (p. 82), “3.6
Changing the Number of Display Digits” (p. 48), and “7.8 Setting Output Formats” (p. 92).
USB COM
,
USB KEYBD
,
RS-232C
,
LAN
:
Voltage (units: mV, V)
Measured value range
Measured value
When
+OvrRng
or
−OvrRng
is displayed
At measurement error
100 mV
±
.
E-03
±990.35
+991.35
1 V
±
.
E-03
±9900.0000E+34
+9910.0000E+34
10 V
±
.
E+00
±99.36
+99.36
100 V
±
.
E+00
±990.35
+991.35
1000 V
±
.
E+00
±9900.0000E+34
+9910.0000E+34
Temperature (unit: °C)
Measured value
When
+OvrRng
or
−OvrRng
is displayed
At measurement error
±
.
±9.900E+37
+9.910E+37
USB MEMORY
:
Voltage (units: mV, V)
Measured value
When
+OvrRng
or
−OvrRng
is displayed
At measurement error
±
.
E±0
±9.937
+9.937
Temperature (unit: °C)
Measured value
When
+OvrRng
or
−OvrRng
is displayed
At measurement error
±
.
E+0
±9.90E+37
+9.91E+37
If there are not enough digits in the integer portion, digits with the value of 0 will be added.
Example: If the measured value in the 1000 V range is 1 V, the value will be indicated as
+0001.0000E+00. In the event of the
+OvrRng
or
−OvrRng
display, the value will be ±9.9E+37,
and in the event of a measured value error, the value will be 9.91E+37.
For more information about output when the interface is set to
PRINTER
, see “Print examples” (p. 148).
www.