157
Functional Specifications
13.3
Functional Specifications
Display measured values
Settings
V, V°C
Default setting
V
Range switching
Settings
AUTO, MANUAL
Default setting
AUTO
Input resistance
switching
Settings
10 M
Ω
, AUTO
(100 V range and 1000 V range: fixed at 10 M
Ω
)
Default setting
10 M
Ω
Display digit selection
Settings
7 1/2 digits, 6 1/2 digits, 5 1/2 digits, 4 1/2 digits, 3 1/2 digits
Default setting
7 1/2 digits
Integration time
Setting
Integration time unit: PLC, ms
PLC setting range: 0.02, 0.2, 1, 10, 100
ms setting range: 1 ms to 9999 ms
Preset integration
times
FAST: 1 PLC
MEDIUM: 10 PLC
SLOW: 100 PLC
Default setting
MEDIUM (10 PLC)
Smoothing function
Operation
Displays the moving average of measured values in the RUN
state.
1
1
n A
smooth
k
k n
V
V
A
+ −
=
=
∑
V
smooth
: Average value
A
: Number of averaging iterations
n
: Number of measurements
V
k
:
k
th
measured value
Settings
Smoothing: ON, OFF
Number of averaging iterations: 2 to 100
Default settings
Smoothing: OFF; number of averaging iterations: 4
Triggers Continuous
measurement
Settings
RUN, STOP
When set to STOP, single trigger from
[TRIG]
key
Default setting
RUN
Trigger source
Setting
INTERNAL, EXTERNAL
When using the EXTERNAL setting, TRIG signal input and
[TRIG]
key input are each treated as one trigger event.
Default setting
INTERNAL
Number of
measurements
Setting
1 per trigger to 5000 per trigger
(Disabled when in the RUN state)
Default setting
1 per trigger
Delay
Settings
Delay: PRESET, MANUAL
PRESET time: 0 ms
MANUAL time: 0 ms to 9999 ms
Default settings
Delay: PRESET; MANUAL time: 0 ms
NULL
Calculation formula
V
M
=
V
-
V
N
V
M
: Measured value after NULL calculation
V
: Voltage measured value
V
N
: NULL value
Settings
NULL: ON, OFF
NULL value: -1000 V to +1000 V (non-range-dependent
value, acquired from current measured value or set as
desired)
Default settings
NULL: OFF; NULL value: 0 V
13